Retrieval Analysis of a Polycarbonate-Urethane Acetabular Cup: A Case Report

Volume 18, Edição 1, 2008, pp. 69-74
DOI: 10.1615/JLongTermEffMedImplants.v18.i1.570
Get accessGet access

RESUMO

An acetabular cup made of a compliant polycarbonate-urethane material has been introduced as an alternative bearing solution. This case study examines the use of this acetabular cup in a single patient at 12 months. The cup had minimal loss of thickness with the most being in the superior area (approximately 10%) and minimal loss of weight (2.4%). On the back side was evidence of abrasive macroscopic wear on one side of the implanted component in the area of directional loading from the head to the acetabulum. On the front side, the wear rate was determined to be minimal (less than 15 mm3 per year), meaning that most of the change in thickness and weight was caused on the back side. The retrieved synovial fluid appeared normal in color and volume at the time of revision. The histology of the tissue taken showed minimal wear particles and minimal reactivity, confirming that the patient did not have any signs of synovitis.The analysis of the cup confirms the preclinically determined low wear articulation and biocompatibility of polycarbonate-urethane as a weight-bearing material. In summary, the surgical findings, data reviewed, and images taken from this case report warrant further study.

CITADO POR
  1. Vani Kohila, Thomas Susha, Prabhawathi V., Boobalan T., Sawant Shilpa N., Doble Mukesh, In vitro biocompatiblity of modified polycarbonate as a biomaterial, Colloids and Surfaces B: Biointerfaces, 108, 2013. Crossref

  2. St. John Kenneth R., The use of polyurethane materials in the surgery of the spine: a review, The Spine Journal, 14, 12, 2014. Crossref

  3. St. John Kenneth R., The use of compliant layer prosthetic components in orthopedic joint repair and replacement: A review, Journal of Biomedical Materials Research Part B: Applied Biomaterials, 102, 6, 2014. Crossref

  4. Gabarre Sergio, Herrera Antonio, Mateo Jesús, Ibarz Elena, Lobo-Escolar Antonio, Gracia Luis, Study of the Polycarbonate-Urethane/Metal Contact in Different Positions during Gait Cycle, BioMed Research International, 2014, 2014. Crossref

  5. Kurtz Steven M., Ong Kevin, Contemporary Total Hip Arthroplasty: Alternative Bearings, in UHMWPE Biomaterials Handbook, 2016. Crossref

  6. Miller Andrew T., Safranski David L., Smith Kathryn E., Guldberg Robert E., Gall Ken, Compressive cyclic ratcheting and fatigue of synthetic, soft biomedical polymers in solution, Journal of the Mechanical Behavior of Biomedical Materials, 54, 2016. Crossref

  7. Sheikhhassani Ramtin, Anvari Pasha, Taei Simin, Sheikhhassani Yasmin, Potential use of a polycarbonate-urethane matrix reinforced with polyethylene fibers for shock-absorbing dental implants, Medical Hypotheses, 85, 3, 2015. Crossref

  8. Alves Diana, Sileika Tadas, Messersmith Phillip B., Pereira Maria Olívia, Polydopamine-Mediated Immobilization of Alginate Lyase to PreventP. aeruginosaAdhesion, Macromolecular Bioscience, 16, 9, 2016. Crossref

  9. Biant Leela C, Gascoyne Trevor C, Bohm Eric R, Moran Matthew, Retrieval analysis of a failed TriboFit polycarbonate urethane acetabular buffer, Proceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 230, 3, 2016. Crossref

  10. St. John Kenneth, Gupta Minakshi, Evaluation of the wear performance of a polycarbonate-urethane acetabular component in a hip joint simulator and comparison with UHMWPE and cross-linked UHMWPE, Journal of Biomaterials Applications, 27, 1, 2012. Crossref

  11. Grieco Preston W., Pascal Scott, Newman Jared M., Shah Neil V., Stroud Sarah G., Sheth Neil P., Maheshwari Aditya V., New alternate bearing surfaces in total hip arthroplasty: A review of the current literature, Journal of Clinical Orthopaedics and Trauma, 9, 1, 2018. Crossref

  12. Elsner Jonathan J., McKeon Brian P., Orthopedic Application of Polycarbonate Urethanes: A Review, Techniques in Orthopaedics, 32, 3, 2017. Crossref

  13. Van Susante Job L C, Verdonschot Nico, Bom L Paul A, Tomaszewski Pawel, Campbell Pat, Ebramzadeh Edward, Schreurs B Wim, Lessons learnt from early failure of a patient trial with a polymer-on-polymer resurfacing hip arthroplasty, Acta Orthopaedica, 89, 1, 2018. Crossref

  14. Ford Audrey C., Gramling Hannah, Li Samuel C., Sov Jessica V., Srinivasan Amrita, Pruitt Lisa A., Micromechanisms of fatigue crack growth in polycarbonate polyurethane: Time dependent and hydration effects, Journal of the Mechanical Behavior of Biomedical Materials, 79, 2018. Crossref

  15. Kanca Yusuf, Milner Piers, Dini Daniele, Amis Andrew A., Tribological evaluation of biomedical polycarbonate urethanes against articular cartilage, Journal of the Mechanical Behavior of Biomedical Materials, 82, 2018. Crossref

  16. Zou Jie, Wang Weiwei, Kratz Karl, Xu Xun, Nie Yan, Ma Nan, Lendlein Andreas, Park J.-W., Krüger A., Jung F., Evaluation of human mesenchymal stem cell senescence, differentiation and secretion behavior cultured on polycarbonate cell culture inserts, Clinical Hemorheology and Microcirculation, 70, 4, 2019. Crossref

  17. Mai Sabine, Mai Burkhard, Siebert Werner E., The Use of Polycarbonate-Urethane as an Acetabular Shell Bearing Surface: A 5-Year Prospective Study, HIP International, 27, 5, 2017. Crossref

  18. Lazic Stefan, Kellett Catherine, Afzal Irrum, Mohan Rajan, Killampalli Vijay, Field Richard E, Three-year results of a polycarbonate urethane acetabular bearing in total hip arthroplasty, HIP International, 30, 3, 2020. Crossref

  19. Moroni Antonio, Hoque Martha, Micera Giovanni, Orsini Riccardo, Nocco Emanuele, Giannini Sandro, Cushion Form Bearings in Total Hip Arthroplasty: Nature’s Approach to the Synovial Joint Problem, in Tribology in Total Hip Arthroplasty, 2011. Crossref

  20. Rajpura A., Kendoff D., Board T. N., The current state of bearing surfaces in total hip replacement, The Bone & Joint Journal, 96-B, 2, 2014. Crossref

  21. Barber Helena, Kelly Cambre N., Abar Bijan, Allen Nicholas, Adams Samuel B., Gall Ken, Rotational Wear and Friction of Ti-6Al-4V and CoCrMo against Polyethylene and Polycarbonate Urethane, Biotribology, 26, 2021. Crossref

  22. Ford Audrey, Hua Zikai, Ferguson Stephen J., Pruitt Lisa A., Gao Leiming, A 3D-transient elastohydrodynamic lubrication hip implant model to compare ultra high molecular weight polyethylene with more compliant polycarbonate polyurethane acetabular cups, Journal of the Mechanical Behavior of Biomedical Materials, 119, 2021. Crossref

  23. Molinari Giampaolo, Galmarini Valter, Ferrara Fabrizio, Capelli Roberto, Treatment of hip fractures in the elderly with the “TriboFit” System: retrospective review of our experience, Lo Scalpello - Otodi Educational, 35, 2, 2021. Crossref

  24. Glatzeder Korbinian, Komnik Igor, Ambellan Felix, Zachow Stefan, Potthast Wolfgang, Dynamic pressure analysis of novel interpositional knee spacer implants in 3D-printed human knee models, Scientific Reports, 12, 1, 2022. Crossref

693 Visualizações do artigo 20 downloads de artigos Métricas
693 VISUALIZAÇÕES 20 TRANSFERÊNCIAS 24 Crossref CITAÇÕES Google
Scholar
CITAÇÕES

Artigos com conteúdo semelhante:

Can Ultrastructural Particle Location Predict Aseptic Loosening? A Biopsy Study of Nonloose Hip Implants One Year Postoperative Using Three Bearing Material Combinations Journal of Long-Term Effects of Medical Implants, Vol.17, 2007, issue 4
F. Elling, Marianne Nygaard, Kjeld Soballe, A. Borgwardt, L. Bastholm
Explantation and Analysis of the First Retrieved Human Acetabular Cup Made of Polycarbonate Urethane: A Case Report Journal of Long-Term Effects of Medical Implants, Vol.18, 2008, issue 1
Richard Treharne, N. J. Hallab, B. Wippermann, Steven M. Kurtz
A Systematic Review Investigating the Early Rehabilitation of Patients Following Medial Patellofemoral Ligament Reconstruction for Patellar Instability Critical Reviews™ in Physical and Rehabilitation Medicine, Vol.19, 2007, issue 2
James Walker, Toby O. Smith, Nicola Russell
C-Reactive Protein and Erythrocyte Sedimentation Rates after Total and Unicompartmental Knee Arthroplasty−Less Implant Equals Quicker Normalization Journal of Long-Term Effects of Medical Implants, Vol.34, 2024, issue 4
Şahin Çepni, Ömer Faruk Naldöven, Kasım Kılıçarslan, Ahmet Fırat, Şahan Güven, Enejd Veizi
10-Year Evaluation of the Cementless Low-Contact- Stress Rotating-Platform Total Knee Arthroplasty Journal of Long-Term Effects of Medical Implants, Vol.19, 2009, issue 4
Vasileios S. Nikolaou, Demetrios S. Korres, Olga D. Savvidou, Panayiotis J. Papagelopoulos, Stergios N. Lallos, Nikolaos E. Efstathopoulos, Andreas F. Mavrogenis

Última edição

Design and Stability Analysis of an Adaptive Neuro-Fuzzy Inference System (ANFIS)−Based Pacemaker Controller in MATLAB Simulink Asghar Dabiri Aghdam, Nader Jafarnia Dabanloo, Fereidoun Nooshiravan Rahatabad, Keivan Maghooli Complications Following Intraosseous Injections of Calcium Phosphate Bone Cement in Subchondroplasty Christos B. Zachariadis, Helen C. Leligou, Stauros K. Kourkoulis, Evaggelos Magnisalis, Panayiotis J. Papagelopoulos, Olga D. Savvidou Complications Following Short Femoral Nail Fixation for Intertrochanteric Hip Fractures: A Retrospective Study Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, C. Nikakis, Stavros Angelis, Spyridon J. Maris, C. Grekas, Efthymios J. Karadimas, Alexandros P. Apostolopoulos Finite Element Analysis: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Ali Muhsin, Enas Kareem Mohammed, Khalid Bander C-Reactive Protein and Erythrocyte Sedimentation Rates after Total and Unicompartmental Knee Arthroplasty−Less Implant Equals Quicker Normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat Comparison of Roughness, Wettability, and SEM Features between Sandblasted Acid-Etched and Oxidized Titanium Dental Implants Kshitiz Chhabra, Arvina Rajasekar Correlation between Primary Stability of Dental Implants and Bone Density: A Retrospective Analysis S. Padmaja, Arvina Rajasekar

Próximos artigos

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka
Portal Digital Begell Biblioteca digital da Begell eBooks Diários Referências e Anais Coleções de pesquisa Políticas de preços e assinaturas Begell House Contato Language English 中文 Русский Português German French Spain