Specific Shortcomings of Endograft Design

Tieying Yin
Dept. of Surgery and Radiology, Laval University and Quebec Biomaterials Institute, CHUQ, Canada; Key Laboratory of Biorheological Science and Technology, Bioengineering College, Chongqing University, Chongqing, PR China
Robert Guidoin
Department of Surgery, Faculty of Medicine, Laval University, Quebec, QC, Canada; Axe Medecine Regeneratrice, Centre de Recherche du CHU, Quebec, QC, Canada
Marc M. Corriveau
Division of Vascular Surgery, McGill University, Montreal
Mark Nutley
Division of Vascular Surgery and Department of Diagnostic Imaging, University of Calgary, Peter Lougheed Health Centre (AB), Calgary, Canada
Lu Xu
Key Laboratory of Textile Science and Technology of Ministry of Education, Donghua University, Shanghai
Georgi R. Marinov
Department of Anatomy, Histology and Embryology, Medical University of Varna, Varna
Lu Wang
Key Laboratory of Textile Science and Technology of the Ministry of Education, College of Textile, Donghua University, Shanghai, China
Yahye Merhi
Laboratory of Thrombosis and Haemostasis Research Centre, Montreal Heart Institute, Montreal, QC, Canada
Robert McGregor
Imaging Department, Boundary Trails Health Centre Winkler, Manitoba, Canada
Ze Zhang
Department of Surgery, Faculty of Medicine, Laval University, Quebec, QC, Canada; Axe Medecine Regeneratrice, Centre de Recherche du CHU, Quebec, QC, Canada
Yvan Douville
Departments of Surgery and Radiology, Laval University and Quebec Biomaterials Institute, CHUQ, Quebec
Stephane Turgeon
Departments of Surgery and Radiology, Laval University and Quebec Biomaterials Institute, CHUQ, Quebec
Martin King
College of Textiles, North Carolina State University, Raleigh, NC
Oren K. Steinmetz
Division of Vascular Surgery, McGill University, Montreal

RESUMO

Further to the rapid enlargement of an aneurysm to 5.6 cm in diameter after 3 years of surveillance, a 79-year-old patient was fitted with a Vanguard modular stent graft and monitored on a regular basis for 6 years. Two years later, the aneurysmal sac ruptured. The patient died 1 month after an open surgery. The device was devoid of any encapsulation and the ipsilateral limb was detached from the body. The Nitinol skeleton was mostly maintained, however, some polypropylene sutures were broken. The resulting motion of the sharp-angled Nitinol wires caused abrasion and resulted in a few localized holes that were sufficient to permit blood to percolate through the textile wall. Some polyester yarns in the warp direction were ruptured. The Nitinol wire used in this device was shown to be corrosion resistant but the selection of the polypropylene suture was inappropriate. Because this technology is maturing rapidly, these weaknesses can be avoided in the future generations of endovascular devices. It is recommended that these Nitinol wires be sutured to the fabric and that polyester yarns stronger than 68 decitex in tubes 8 mm in diameter are selected.


Articles with similar content:

Persistent Type II Endoleak Unrelated to an Anaconda Aortic Stent Graft Fulfilling the 3Bs Requirements of Biofunctionality, Biodurability, and Biocompatibility
Journal of Long-Term Effects of Medical Implants, Vol.18, 2008, issue 3
Xingyi Xie, Robert Guidoin, Randolph Guzman, Tieying Yin, Robert McGregor, Yvan Douville, Martin King, Mahmoud Rouabhia, Georgi R. Marinov, Mark Nutley, Lu Xu, Yahye Merhi, Ze Zhang, Lu Wang, Guixue Wang, Guy Dionne, Stephane Turgeon
Characterization of an Endovascular Prosthesis Using the 3Bs Rule (Biocompatibility, Biofunctionality and Biodurability): A Recommended Protocol to Investigate a Device Harvested at Necropsy
Journal of Long-Term Effects of Medical Implants, Vol.17, 2007, issue 3
Robert Guidoin, Xiaoyan Deng, Randolph Guzman, Tieying Yin, Robert McGregor, Yvan Douville, Martin King, Jean-Pierre Renou, Georgi R. Marinov, Hongbo Zhang, Yubo Fan, Robert M. Geelkerken, Mark Nutley, Ze Zhang, Zaipin Xu, Andre-Pierre Legrand, Guy Dionne, Stephane Turgeon, Nilesh Ingle, Amidou Traore
Detailed Analysis of a Series of Explanted Talent AAA Stent-Grafts: Biofunctionality Assessment
Journal of Long-Term Effects of Medical Implants, Vol.21, 2011, issue 4
Nathalie Gilbert, Robert Guidoin, Tieying Yin, Robert McGregor, Yvan Douville, Gregory Samis, Georgi R. Marinov, Jing Lin, Benedikt Weber, Mark Nutley, Yahye Merhi, Ze Zhang, Dangheng Wei, Lu Wang, Bin Li, Guy Dionne
Fatigue and/or Failure Phenomena Observed in the Fabric of Stent-Grafts Explanted After Adverse Events
Journal of Long-Term Effects of Medical Implants, Vol.23, 2013, issue 1
Nathalie Gilbert, Thien How, Robert Guidoin, Royston Paynter, Yvan Douville, Gregory Samis, Helene Crepeau, Jing Lin, Mark Nutley, Ze Zhang, Dangheng Wei, Lu Wang, Guy Dionne
Clinical Evaluation of a New Total Hip Prosthetic Design: 100 Consecutive Cementless Total Hip Arthroplasties Using Sulzermedica's "Natural Hip" With Two- to Six-Year Clinical and Radiographic Follow-Up
Critical Reviews™ in Biomedical Engineering, Vol.28, 2000, issue 1&2
Thomas B. Pace