ライブラリ登録: Guest
Critical Reviews™ in Biomedical Engineering

年間 6 号発行

ISSN 印刷: 0278-940X

ISSN オンライン: 1943-619X

SJR: 0.262 SNIP: 0.372 CiteScore™:: 2.2 H-Index: 56

Indexed in

Measurements of Enlarged Blood Pump Models Using Laser Doppler Anemometer

巻 28, 発行 1&2, 2000, pp. 135-140
DOI: 10.1615/CritRevBiomedEng.v28.i12.230
Get accessGet access

要約

In an earlier study (Chua et al., 1998, 1999a), a 5:1 enlarged model of the Kyoto-NTN Magnetically Suspended Centrifugal Blood Pump (Akamatsu et al., 1995) with five different impeller blade profiles was designed and constructed. Their respective flow characteristics with respect to (1) the three different blade profile designs: forward, radial, and backward, (2) the number of blades used, and (3) the rotating speed were investigated. Among the five impeller designs, the results obtained suggested that impellers A and С designs should be adopted if higher head is required. Impellers A and С therefore were selected for the flow in between their blades to be measured using Laser Doppler Anemometer (LDA), so as to have a better understanding of the flow physics with respect to the design parameters.

Begell Digital Portal Begellデジタルライブラリー 電子書籍 ジャーナル 参考文献と会報 リサーチ集 価格及び購読のポリシー Begell House 連絡先 Language English 中文 Русский Português German French Spain