Volume 51, 2020 Volume 50, 2019 Volume 49, 2018 Volume 48, 2017 Volume 47, 2016 Volume 46, 2015 Volume 45, 2014 Volume 44, 2013 Volume 43, 2012 Volume 42, 2011 Volume 41, 2010 Volume 40, 2009 Volume 39, 2008 Volume 38, 2007 Volume 37, 2006 Volume 36, 2005 Volume 35, 2004 Volume 34, 2003 Volume 33, 2002 Volume 32, 2001 Volume 31, 2000 Volume 30, 1999 Volume 29, 1998 Volume 28, 1997

Heat Transfer Research

DOI: 10.1615/HeatTransRes.2017019111
pages 1421-1432

A TEMPERATURE ERROR CORRECTION METHOD FOR A DTR503A NATURALLY VENTILATED RADIATION SHIELD

Jie Yang
Jiangsu Collaborative Innovation Center on Atmospheric Environment and Equipment Technology, Nanjing 210044, China; Jiangsu Key Laboratory of Meteorological Observation and Information Processing, Nanjing 210044, China
Qingquan Liu
Jiangsu Collaborative Innovation Center on Atmospheric Environment and Equipment Technology, Nanjing 210044, China; Jiangsu Key Laboratory of Meteorological Observation and Information Processing, Nanjing 210044, China
Wei Dai
Key Laboratory of MEMS of the Ministry of Education, Nanjing 210009, China

RÉSUMÉ

Due to solar radiation exposure, air flowing inside a naturally ventilated radiation shield may produce a measurement error of 0.8°C or higher. To improve the air temperature observation accuracy, a temperature error correction method is proposed. The correction method is based on a computational fluid dynamics (CFD) method and a genetic algorithm (GA) method. The CFD method is implemented to obtain the temperature error of a DTR503A naturally ventilated radiation shield under various environmental conditions. Then, a temperature error correction equation is obtained by fitting the CFD results using the GA method. To verify the performance of the correction equation, the DTR503A naturally ventilated radiation shield and an aspirated temperature measurement platform are characterized in the same environment to allow intercomparison. The aspirated temperature measurement platform serves as an air temperature reference. The mean temperature error given by the intercomparison experiments is 0.66°C, and the mean temperature error given by the correction equation is 0.67°C. This correction equation allows the temperature error to be reduced by approximately 98.5%. The mean absolute error and the root mean square error between the temperature errors given by the correction equation and the temperature errors given by the experiments are 0.073°C and 0.079°C, respectively. The correction equation is only for the DTR503A naturally ventilated radiation shield, but the temperature error correction method can be used on a variety of shields.