Effect of Implant-Abutment Connection Type on Stress Distribution in Peri-Implant Bone and Abutment Micromovement: A Three-Dimensional Finite Element Analysis

Volume 29, Numéro 2, 2019, pp. 113-124
DOI: 10.1615/JLongTermEffMedImplants.2019031391
Get accessGet access

RÉSUMÉ

This study assessed the effect of implant-abutment connection type on stress distribution in peri-implant bone and abutment micromovement using finite-element analysis (FEA). Dimensions of three implant-abutment designs were measured by a three-dimensional scanner and transferred to SOLIDWORKS. An elemental model was designed using ABAQUS. Each implant was placed in bone at the crestal level. A 100-N load was applied at a 45° angle relative to the longitudinal axis of the implant in the buccolingual direction from the palatal toward the buccal and coronal parts of the abutment. Shear stress, von Mises stress, and micromovement of the abutment relative to the fixture were analyzed. Data were reported qualitatively and quantitatively using ABAQUS. The von Mises stress in all three samples decreased from the crest toward the apical area and was distributed homogenously. Maximum stress concentration was at the most coronal part of the implant body in the midbuccal area. In cancellous and cortical bone, stress decreased from the crest toward the palate and was at its maximum at the midbuccal point in the bone crest. Stress in cortical bone was more homogenous and in cancellous bone was higher. Shear stress was higher in the buccal than in the palatal area, and at its maximum shear stress and equal in the mesiobuccal and distobuccal areas. Micromovement was 4.25 μm in an Astra implant, 5.42 μm in a Intra-Lock implant, and 6.63 μm in an SPI implant. The distribution of von Mises and shear stress was the same in bone around the three implant connection types; however, abutment micomovement differed.

RÉFÉRENCES
  1. Yamanishi YYS, Imazato S, Nakano T, Yatani H. Influences of implant neck design and implant-abutment joint type on peri-implant bone stress and abutment micromovement: three-dimensional finite element analysis. Dent Mater. 2012;28(11):1126-33. .

  2. Albrektsson T, Zarb G, Worthington P, Eriksson A. The long-term efficacy of currently used dental implants: a review and proposed criteria of success. Int J Oral Maxillofac Surg. 1986;1(1):11-25. .

  3. Oh T-J, Yoon J, Misch CE, Wang H-L. The causes of early implant bone loss: myth or science. J Periodontol. 2002;73(3):322-33. .

  4. Lee DW, Choi YS, Park KH, Kim CS, Moon IS. Effect of microthread on the maintenance of marginal bone level: a 3-year prospective study. Clin Oral Implants Res. 2007;18(4):465-70 .

  5. Becker W, Goldstein M, Becker BE, Sennerby L. Minimally invasive flapless implant surgery: a prospective multicenter study. Clin Implant Dentistry Related Res. 2005;7:s21-7. .

  6. Kim Y, Oh TJ, Misch CE, Wang HL. Occlusal considerations in implant therapy: clinical guidelines with biomechanical rationale. Clin Oral Implants Res. 2005;16(1): 26-35. .

  7. King GN, Hermann JS, Schoolfield JD, Buser D, Cochran DL. Influence of the size of the microgap on crestal bone levels in non-submerged dental implants: a radiographic study in the canine mandible. J Periodontol. 2002;73(10):1111-7. .

  8. Weng D, Nagata M, Bell M, de Melo L, Bosco AF. Influence of microgap location and configuration on peri-implant bone morphology in nonsubmerged implants: an experimental study in dogs. Int J Oral Maxillofac Implants. 2010;25(3):540-7. .

  9. Verborgt O, Gibson GJ, Schaffler MB. Loss of osteocyte integrity in association with microdamage and bone remodeling after fatigue in vivo. J Bone Miner Res. 2000;15(1):60-7. .

  10. Huiskes R, Ruimerman R, Van Lenthe GH, Janssen JD. Effects of mechanical forces on maintenance and adaptation of form in trabecular bone. Nature. 2000;405(6787):704-6. .

  11. Holmgren EP, Seckinger RJ, Kilgren LM, Mante F. Evaluating parameters of osseointegrated dental implants using finite element analysis-a two-dimensional comparative study examining the effects of implant diameter, implant shape, and load direction. J Oral Implantol. 1998;24(2): 80-8. .

  12. Assif D, Marshak B, Horowitz A. Analysis of load transfer and stress distribution by an implant-supported fixed partial denture. J Prosth Dent. 1996;75(3):285-91. .

  13. Schrotenboer J, Tsao Y-P, Kinariwala V, Wang H-L. Effect of microthreads and platform switching on crestal bone stress levels: a finite element analysis. J Periodontol. 2008;79(11):2166-72. .

  14. Schrotenboer J, Tsao Y-P, Kinariwala V, Wang H-L. Effect of platform switching on implant crest bone stress: a finite element analysis. Implant Dent. 2009;18(3):260-9. .

  15. Hudieb MI, Wakabayashi N, Kasugai S. Magnitude and direction of mechanical stress at the osseointegrated interface of the microthread implant. J Periodontol. 2011;82(7):1061-70. .

  16. Chang C-L, Chen C-S, Hsu ML. Biomechanical effect of platform switching in implant dentistry: a three-dimensional finite element analysis. Int J Oral Maxillofac Implants. 2010;25(2):295-304 .

  17. Hsu J-T, Fuh L-J, Lin D-J, Shen Y-W, Huang H-L. Bone strain and interfacial sliding analyses of platform switching and implant diameter on an immediately loaded implant: experimental and three-dimensional finite element analyses. J Periodontol. 2009;80(7):1125-32. .

  18. Park IS, Won SY, Bae TS, Song KY, Park CW, Eom TG, Jeong CM. Fatigue characteristics of five types of implant-abutment joint designs. Metals Mater Int. 2008;14(2): 133-8. .

  19. Harder S, Dimaczek B, Afil Y, Terheyden H, Freitag-Wolf S, Kern M. Molecular leakage at implant-abutment connection-in vitro investigation of tightness of internal conical implant-abutment connections against endotoxin penetration. Clin Oral Investig. 2010;14(4):427-32. .

  20. Tsuge T, Hagiwara Y, Matsumura H. Marginal fit and microgaps of implant-abutment interface with internal anti-rotation configuration. Dent Mater J. 2008;27(1):29-34. .

  21. Rack A, Rack T, Stiller M, Riesemeier H, Zabler S, Nelson K. In vitro synchrotron-based radiography of micro-gap formation at the implant-abutment interface of two-piece dental implants. J Synchrotron Radiat. 2010;17(2):289-94. .

  22. Proff P, Steinmetz I, Bayerlein T, Dietze S, Fanghanel J, Gedrange T. Bacterial colonisation of interior implant threads with and without sealing. Folia Morphol (Warsz). 2006;65(1):75-7. .

  23. Myshin HL, Wiens JP. Factors affecting soft tissue around dental implants: a review of the literature. J Prosthet Dent. 2005;94(5):440-4. .

  24. Faggion CM, Listl S, Tu Y-K. Assessment of endpoints in studies on peri-implantitis treatment-a systematic review. J Dent. 2010;38(6):443-50. .

  25. Steinebrunner L, Wolfart S, BoBmann K, Kern M. In vitro evaluation of bacterial leakage along the implant-abutment interface of different implant systems. Int J Oral Maxillofac Implants. 2005;20(6); 875-81. .

  26. Shareef N, Levine D. Effect of manufacturing tolerances on the micromotion at the Morse taper interface in modular hip implants using the finite element technique. Biomaterials. 1996;17(6):623-30. .

  27. Segundo R, Oshima H, da Silva I, Burnett L Jr, Mota EG, Silva LL. Stress distribution of an internal connection implant prostheses set: A 3D finite element analysis. Stomatologija. 2009;11(2):55-9. .

  28. Segundo R, Oshima H, Silva I, Junior L, Mota EG, Coelho L. Stress distribution on external hexagon implant system using 3D finite element analysis. Acta Odontologica Lati31. Broggini N, McManus L, Hermann J, Medina R, Oatesnoamericana. 2007;20(2):79-81. .

  29. Saidin S, Kadir MRA, Sulaiman E, Kasim NHA. Effects of different implant-abutment connections on micromotion and stress distribution: prediction of microgap formation. J Dent. 2012;40(6):467-74. .

  30. Patterson EA, Johns R. Theoretical analysis of the fatigue life of fixture screws in osseointegrated dental implants. Int J Oral Maxillofac Implants. 1992;7(1):26-33. .

  31. Broggini N, McManus L, Hermann J, Medina R, Oates T, Schenk R, Buser D, Mellonig J, Cochran D. Persistent acute inflammation at the implant-abutment interface. J Dent Res. 2003;82(3):232-7. .

  32. Bozkaya D, Muftu S. Mechanics of the tapered interference fit in dental implants. J Biomech. 2003;36(11):1649-58. .

  33. Dibart S, Warbington M, Su MF, Skobe Z. In vitro evaluation of the implant-abutment bacterial seal: the locking taper system. Int J Oral Maxillofac Implants. 2005;20(5):732-7. .

  34. Zipprich H, Weigl P, Lange B, Lauer H. Micromovements at the implant-abutment interface: Measurement, causes, and consequences. Implantologie. 2009;15:31-45 (in German). .

  35. Abdul-Kadir MR, Hansen U, Klabunde R, Lucas D, Amis A. Finite element modelling of primary hip stem stability: the effect of interference fit. J Biomech. 2008;41(3):587-94. .

  36. Semper W, Kraft S, Kruger T, Nelson K. Theoretical considerations: implant positional index design. J Dent Res. 2009;88(8):725-30. .

  37. Huang H-L, Chang C-H, Hsu J-T, Faligatter AM, Ko CC. Comparison of implant body designs and threaded designs of dental implants: a 3-dimensional finite element analysis. Int J Oral Maxillofac Implants. 2007;22(4):551-62. .

  38. Hsu M-L, Chang CL. Application of finite element analysis in dentistry. In: Moratal D, editor. Finite element analysis. London: InTech Open; 2010. p. 43-60. .

  39. Rahimi A, Bourauel C, Jager A, Gedrange T, Heinemann F. Load transfer by fine threading the implant neck-a FEM study. J Physiol Pharmacol. 2009;60(Suppl 8):107-12. .

  40. Ahlberg JP, Kovero OA, Hurmerinta KA, Zepa I, Nissinen MJ, Kononen MH. Maximal bite force and its association with signs and symptoms of TMD, occlusion, and body mass index in a cohort of young adults. Cranio. 2003;21(4):248-52. .

  41. Koc D, Dogan A, Bek B. Bite force and influential factors on bite force measurements: a literature review. Eur J Dent. 2010;4(2):223-32. .

  42. Hansson S. A conical implant-abutment interface at the level of the marginal bone improves the distribution of stresses in the supporting bone. Clini Oral Implants Res. 2003;14(3):286-93. .

  43. Misch CE. Bone density: a key determinant for treatment planning. Contemp Implant Dent. St. Louis, MO: Mosby; 2008. p. 130-46. .

  44. Guo X. Mechanical properties of cortical bone and cancellous bone tissue. Bone. 2001;10:1-23. .

  45. Norton M. An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design. Clin Oral Implants Res. 1997;8(4):290-8. .

  46. Kayabasi O, Yuzbasioglu E, Erzincanli F. Static, dynamic and fatigue behaviors of dental implant using finite element method. Adv Eng Software. 2006;37(10):649-58. .

  47. Huang H-L, Chang C-H, Hsu J-T, Faligatter AM, Ko C-C. Comparison of implant body designs and threaded designs of dental implants: a 3-dimensional finite element analysis. Int J Oral Maxillofac Implants. 2007;22(4). .

CITÉ PAR
  1. Lorusso Felice, Tartaglia Gianluca, Inchingolo Francesco, Scarano Antonio, Early Response and Clinical Efficacy of a Mouthwash Containing Chlorhexidine, Anti Discoloration System, Polyvinylpyrrolidone/Vinyl Acetate and Sodium DNA in Periodontitis Model: A Triple-Blind Randomized Controlled Clinical Trial, Dentistry Journal, 10, 6, 2022. Crossref

  2. Scarano Antonio, Tartaglia Gianluca, Rapone Biagio, Inchingolo Francesco, Lorusso Felice, Maxillary Sinus Osteoma as a Support for Dental Implant Associated to Sinus Augmentation Procedure: A Case Report and Literature Review, Applied Sciences, 12, 11, 2022. Crossref

  3. de Matos Jefferson David Melo, Gomes Leonardo Silva, de Carvalho Ramos Nathália, Queiroz Daher Antonio, Tribst João Paulo Mendes, Campos Tiago Moreira Bastos, Borges Alexandre Luiz Souto, da Rocha Scalzer Lopes Guilherme, Bottino Marco Antonio, Paes Junior Tarcisio José Arruda, Influence of CAD/CAM Abutment Heights on the Biomechanical Behavior of Zirconia Single Crowns, Metals, 12, 12, 2022. Crossref

1517 Vues d'articles 36 Téléchargements d'articles Métrique
1517 VUES 36 TÉLÉCHARGEMENTS 3 Crossref CITATIONS Google
Scholar
CITATIONS

Articles avec un contenu similaire:

Effect of Implant Design on Stress Distribution: A Finite Element Study Journal of Long-Term Effects of Medical Implants, Vol.32, 2022, issue 4
Mohammadreza Talebi Ardakani, Behzad Houshmand, Iman Zoljanahi Oskoui, Anahita Moscowchi
Maximum Insertion Torque, Stress, and Strain in Bone during Insertion of Three Dental Implants with Different Macroscopic Designs: A Dynamic Finite-Element Analysis Journal of Long-Term Effects of Medical Implants, Vol.31, 2021, issue 2
Shirin Zahra Farhad, Majid Torabzadeh, Reza Amid
Static, Dynamic, and Fatigue Finite Element Analysis of Dental Implants with Different Thread Designs Journal of Long-Term Effects of Medical Implants, Vol.26, 2016, issue 4
Hamidreza Katoozian, Mahdi Kadkhodazadeh, Maryam Geramizadeh, Reza Amid
Comparison of Periodontally Compromised Splinted Teeth and Implant Supported Fixed Partial Denture: A Three-Dimensional Finite Element Analysis on Bone Response Journal of Long-Term Effects of Medical Implants, Vol.31, 2021, issue 2
Mohammad Reza Talebi Ardakani, Mohammad Reza Movahhedy, Reza Amid, Mahyar Broukhim, Mahdi Kadkhodazadeh, Arman Hakimi, Mahdieh Mirakhori
BIOMECHANICAL EVALUATION OF MARGINAL BONE LOSS IN THE SURROUNDING BONE UNDER DIFFERENT LOADING: 3D FINITE ELEMENT ANALYSIS STUDY International Journal for Multiscale Computational Engineering, Vol.20, 2022, issue 4
Ali Merdji, Perk Lin Chong, Laid Aminallah, Velaphi Msomi, Abdelhak Ouldyerou, Sandipan Roy

Prochains articles

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar CORRELATION BETWEEN PRIMARY STABILITY OF DENTAL IMPLANTS AND BONE DENSITY: A RETROSPECTIVE ANALYSIS Padmaja S, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos FEA: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Muhsin, Enas Mohammed, Khalid Bander Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka
Portail numérique Bibliothèque numérique eBooks Revues Références et comptes rendus Collections Prix et politiques d'abonnement Begell House Contactez-nous Language English 中文 Русский Português German French Spain