Back-Side Wear in HexLoc Cups Clinico-Radiological, Immunohistopathological, Finite Element, and Retrieval Analysis Studies

Volume 24, Numéro 4, 2014, pp. 319-331
DOI: 10.1615/JLongTermEffMedImplants.2014011940
Get accessGet access

RÉSUMÉ

The HexLoc locking system was designed to prevent back-side wear of the polyethylene liner in the modular cementless metal-backed acetabular cup, but failed. Back-side wear was analyzed using clinico-radiological data, immunohistopathology, finite element modeling (FEM, and retrieval analysis. Screw holes allowed entry of titanium oxide and exit of polyethylene particles. Birefringent polyethylene wear particles were found behind the metal cup in macrophages containing pro-inflammatory tumor necrosis factor-α and interleukin-1β, whereas fibroblast-like cells stained for osteoclastogenic receptor activator of nuclear factor kappa B ligand (RANKL). Computerized tomography revealed granulomas (83% versus 17 %) and cortical destruction (50% versus 5%) better than radiographs. In FEM, a change of the abduction angle from 45 to 60 deg, and liner thickness from 4.8 mm to 2.5 mm, increased the back-side wear by 90% and 120%, respectively. Screw holes were stress concentration areas; their removal decreased wear by 40%. Modeling results were validated in retrieved implants, which demonstrated extensive back-side wear damage of liners with a high abduction angle. Combined clinico-radiological, immunohistopathological, FEM, and retrieval analysis disclosed that back-side wear in the HexLoc design is sensitive to the abduction angle, liner thickness, and presence of screw holes.

CITÉ PAR
  1. Puente Reyna Ana Laura, Jäger Marcus, Floerkemeier Thilo, Frecher Sven, Delank Karl-Stefan, Schilling Christoph, Grupp Thomas M., Backside Wear Analysis of Retrieved Acetabular Liners with a Press-Fit Locking Mechanism in Comparison to Wear Simulation In Vitro, BioMed Research International, 2016, 2016. Crossref

  2. Puente Reyna A. L., Holderied M., Jäger M., Schilling C., Grupp T. M., ARTICULATION AND BACKSIDE WEAR ANALYSIS AFTER LONG-TERM IN VITRO WEAR SIMULATION OF VITAMIN E STABILIZED POLYETHYLENE ACETABULAR LINERS WITH A PRESS-FIT LOCKING MECHANISM, Traumatology and Orthopedics of Russia, 24, 2, 2018. Crossref

  3. Baghdadi Josef, Alkhateeb Shareef, Roth Alexander, Jäger M, Busch A, Alkhateeb S, Landgraeber S, Serong S, Haversath M, vonWasen A, Windhagen H, Flörkemeier T, Budde S, Kubilay J, Noll Y, Delank KS, Baghdadi J, Willburger R, Dücker M, Wilke A, Hütter F, Jäger Marcus, Cup positioning and its effect on polyethylene wear of vitamin E- and non-vitamin E-supplemented liners in total hip arthroplasty: radiographic outcome at 5-year follow-up, Archives of Orthopaedic and Trauma Surgery, 2022. Crossref

Prochains articles

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar CORRELATION BETWEEN PRIMARY STABILITY OF DENTAL IMPLANTS AND BONE DENSITY: A RETROSPECTIVE ANALYSIS Padmaja S, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos FEA: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Muhsin, Enas Mohammed, Khalid Bander Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka
Portail numérique Bibliothèque numérique eBooks Revues Références et comptes rendus Collections Prix et politiques d'abonnement Begell House Contactez-nous Language English 中文 Русский Português German French Spain