Improving the Accuracy of Acetabular Cup Implantation Using a Bulls-Eye Spirit Level

Volumen 21, Edición 3, 2011, pp. 219-224
DOI: 10.1615/JLongTermEffMedImplants.v21.i3.50
Get accessGet access

SINOPSIS

Acetabular introducers have a built-in inclination of 45 degrees to the handle shaft. With patients in the lateral position, surgeons aim to align the introducer shaft vertical to the floor to implant the acetabulum at 45 degrees. We aimed to determine if a bulls-eye spirit level attached to an introducer improved the accuracy of implantation.
A small circular bulls-eye spirit level was attached to the handle of an acetabular introducer. A saw bone hemipelvis was fixed to a horizontal, flat surface. A cement substitute was placed in the acetabulum and subjects were asked to implant a polyethylene cup, aiming to obtain an angle of inclination of 45 degrees. Two attempts were made with the spirit level masked and two with it unmasked. The distance of the air bubble from the spirit level's center was recorded by a single assessor. The angle of inclination of the acetabular component was then calculated. Subjects included both orthopedic consultants and trainees.
Twenty-five subjects completed the study. Accuracy of acetabular implantation when using the unmasked spirit level improved significantly in all grades of surgeon. With the spirit level masked, 12 out of 50 attempts were accurate at 45 degrees inclination; 11 out of 50 attempts were "open," with greater than 45 degrees of inclination, and 27 were "closed," with less than 45 degrees. With the spirit level visible, all subjects achieved an inclination angle of exactly 45 degrees.
A simple device attached to the handle of an acetabular introducer can significantly improve the accuracy of implantation of a cemented cup into a saw bone pelvis in the lateral position.

Próximos Artículos

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar CORRELATION BETWEEN PRIMARY STABILITY OF DENTAL IMPLANTS AND BONE DENSITY: A RETROSPECTIVE ANALYSIS Padmaja S, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos FEA: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Muhsin, Enas Mohammed, Khalid Bander Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka