Manufacturing Imperfection of a Locking Plate Screw Hole: A Case Report

Christopher A. Radkowski
Division of Orthopaedic Surgery, Duke University Medical Center, Durham, NC, USA
Ryan W. Simovitch
Division of Orthopaedic Surgery, Duke University Medical Center, Durham, NC, USA
Robert D. Zura
Assistant Professor of Orthopaedic Surgery, Duke University Medical Center, Division of Orthopaedic Surgery, Box 3389, Durham, NC 27710, USA

SINOPSIS

Plate and screw fixation devices are essential in the orthopaedic surgical management of various fractures. Production of these devices has been nearly perfected, with few manufacturing defects reported. Assuming that all implants have been machined properly may cause a surgeon to overlook a manufacturing defect and may potentially result in a poor patient outcome. We report a case of the imperfect manufacturing of a less invasive stabilization system (LISS) plate screw hole. This resulted in a loss of fixation of a single locking screw to the plate in the treatment of a distal femoral fracture. Prompt recognition of the defect helped avoid a potential adverse outcome for the patient. When these issues arise, open dialogue with industry is crucial to promptly identify a solution and enhance patient care. This case highlights the manufacturing company's outstanding quality control and prompt solution of such an occurrence. It is the purpose of this report to raise surgeons' awareness of potential manufacturing problems with implants and prevent negative consequences as a result.


Articles with similar content:

Prevention of Disabling Back Injuries in Nurses by the Use of Mechanical Patient Lift Systems
Journal of Long-Term Effects of Medical Implants, Vol.14, 2004, issue 6
William B. Long III, Richard Edlich, Kathryne L. Winters, L. D. Britt, Mary Anne Hudson
Reducing Workers' Compensation Costs for Latex Allergy and Litigation against Glove Manufacturing Companies
Journal of Environmental Pathology, Toxicology and Oncology, Vol.28, 2009, issue 4
Jill J. Dahlstrom, William B. Long III, Shelley S. Mason, Richard Edlich, Erin M. Swainston, K. Dean Gubler
Latex Allergy: A Life-Threatening Epidemic for Scientists, Healthcare Personnel, and Their Patients
Journal of Long-Term Effects of Medical Implants, Vol.13, 2003, issue 1
Lisa G. Hill, Richard Edlich, Cynthia L. Heather, Charles R. Woodard
Antibiotic-Loaded Bone Cement and Periprosthetic Joint Infection
Journal of Long-Term Effects of Medical Implants, Vol.24, 2014, issue 2-3
Antonia F. Chen, Javad Parvizi
The Role of Powered Instrumentation in Rhinoplasty and Septoplasty
Journal of Long-Term Effects of Medical Implants, Vol.15, 2005, issue 3
Manuel A. Lopez, John G. Westine, Dean M. Toriumi