Abo Bibliothek: Guest
Digitales Portal Digitale Bibliothek eBooks Zeitschriften Referenzen und Berichte Forschungssammlungen
Critical Reviews™ in Biomedical Engineering
SJR: 0.207 SNIP: 0.376 CiteScore™: 0.79

ISSN Druckformat: 0278-940X
ISSN Online: 1943-619X

Volumes:
Volumen 47, 2019 Volumen 46, 2018 Volumen 45, 2017 Volumen 44, 2016 Volumen 43, 2015 Volumen 42, 2014 Volumen 41, 2013 Volumen 40, 2012 Volumen 39, 2011 Volumen 38, 2010 Volumen 37, 2009 Volumen 36, 2008 Volumen 35, 2007 Volumen 34, 2006 Volumen 33, 2005 Volumen 32, 2004 Volumen 31, 2003 Volumen 30, 2002 Volumen 29, 2001 Volumen 28, 2000 Volumen 27, 1999 Volumen 26, 1998 Volumen 25, 1997 Volumen 24, 1996 Volumen 23, 1995

Critical Reviews™ in Biomedical Engineering

DOI: 10.1615/CritRevBiomedEng.v28.i12.60
pages 23-32

The Biomechanical and Surgical Solutions for Patella Implant in Total Knee Arthroplasty

David G. Mendes
The Center for Implant Surgery, Bnai Zion Medical Center, Haifa, Israel
Mazen Said
The Center for Implant Surgery, Bnai Zion Medical Center, Haifa, Israel
Sorin Head
The Center for Implant Surgery, Bnai Zion Medical Center, Haifa, Israel
Edmond Sabo
The Center for Implant Surgery, Bnai Zion Medical Center, Haifa, Israel
Ricardo Nachman
The Center for Implant Surgery, Bnai Zion Medical Center, Haifa, Israel
Jochanan H. Boss
The Center for Implant Surgery, Bnai Zion Medical Center, Haifa, Israel

ABSTRAKT

A basic fault in designs of patella implants that calls for a prompt remedy is the inadequate polyethylene thickness in contact areas. The requirement for thickness of patella implants should be similar to the requirement set up by the FDA in 1993 regarding the tibial implants, that is, a minimal thickness of 8 mm.
A solution is presented for the optimal design of the patella implant and for the appropriate surgical technique.
The implant's undersurface that fixes to bone is concave and has a 1-cm wide but 5 to 6 mm short central peg that does not risk the integrity of the bone. Fixation of the concave aspect to the convex aspect of the bony patella strongly enhances resistance to medio-lateral and supero-inferior shear forces.
The circumferential facet of the implant 8 mm thick assures uniform thickness of the entire implant and assures optimal mechanical properties of the polyethylene.
The articulating surface is tailored to conform to the corresponding femoral trochlear groove and condyles. It can be made to fit any femoral component of any knee implant.


Articles with similar content:

Long-Term Results of a Cementless Knee Prosthesis with a Metal-Backed Patellar Component: Clinical and Radiological Follow-Up with Histology from Retrieved Components
Journal of Long-Term Effects of Medical Implants, Vol.13, 2003, issue 4
Mikael Sundfeldt, Tomas Albrektsson, Lars V Carlsson, Carina B. Johansson, Lars Regner
Case Report: Pseudotumor associated with corrosion of a femoral component with a modular neck and a ceramic-on-polyethylene bearing
Journal of Long-Term Effects of Medical Implants, Vol.24, 2014, issue 1
Stuart B. Goodman, Joe Messana, Muyibat Adelani
Fixed Versos Mobile Bearing Knee Arthroplasty: A Review of Kinematics and Results
Journal of Long-Term Effects of Medical Implants, Vol.21, 2011, issue 3
A. P. Apostolopoulos, E. Chronopoulos, G. Papachristou, loannis V. Michos, Stergios N. Lallos, Nikolaos E. Efstathopoulos, Andreas F. Mavrogenis
Nontraumatic Fracture of a Custom-Made Wagner Cone Prosthesis Hip Stem: A Case Report
Journal of Long-Term Effects of Medical Implants, Vol.27, 2017, issue 1
Vassilios Nikolaou, Michael-Alexander Malahias, George C. Babis, Evangelos Tyrpenou
Abdominal Wall Reconstruction with Implantable Meshes
Journal of Long-Term Effects of Medical Implants, Vol.21, 2011, issue 1
John M. Felder III, Derek Masden, Christopher E. Attinger, Parag Bhanot, Matthew L. lorio