Stress and Strain Distribution Patterns in Bone around Splinted Standard and Short Implants Placed at the Crestal Level and Subcrestally using Three- Dimensional Finite Element Analysis

Volumen 27, Ausgabe 1, 2017, pp. 1-11
DOI: 10.1615/JLongTermEffMedImplants.2017019926
Get accessGet access

ABSTRAKT

Short implants can be used as alternatives to standard implants to prevent invasive surgical procedures. However, due to concerns about complications caused by less bone–implant contact area, researchers have focused on biomechanical properties of short implants and methods to promote them. Splinting has been suggested to decrease the limitation of short implants. This study compared the pattern of stress and strain distribution in bone supporting splinted standard and short implants positioned at crestal and subcrestal levels. An edentulous posterior mandible was made using computer-aided design. Five models of different combinations of splinted short (4 × 6 mm) and standard (4 × 10 mm) implants placed at the level of crestal bone or subcrestally mesial and distal to the edentulous region with a pontic between them were designed using the CATIA software program. ANSYS software was used for finite element analysis (FEA). In each model, 100 and 300 N loads at zero (parallel to the long axis of implants) and 30° angles were applied to implants. Maximum stress and strain for each of the five models, including equivalent stress, shear stress, and strain in peri-implant cortical and cancellous bone, were calculated and stress distribution pattern in different models were recorded. The highest stress was caused by the 300 N load applied at a 30° angle, followed by the 300 N load applied axially and the 100 N load applied at 30°. This order changed in model 1, where the highest stress was noted under the 300 N load at 30°, followed by the 100 N load at 30°. Maximum stress in peri-implant bone occurred under oblique (30°) load. Maximum stress was noted when two splinted short implants were placed subcrestally. In addition, stress in bone around crestal-level splinted short implants was lower than that around standard implants. Combination of short and standard implants had no biomechanical advantage. Application of load parallel to the long axis can significantly decrease stress in peri-implant bone. Although the combination of short and standard implants has no biomechanical advantage, crestal-level placement of splinted short implants is a suitable treatment plan.

REFERENZIERT VON
  1. Kul Esra, Korkmaz İsmail Hakkı, Effect of different design of abutment and implant on stress distribution in 2 implants and peripheral bone: A finite element analysis study, The Journal of Prosthetic Dentistry, 126, 5, 2021. Crossref

  2. NAZLI Afra Nur, KILINÇ Yeliz, ÇETİNER Sedat, short implants, ADO Klinik Bilimler Dergisi, 2022. Crossref

1493 Artikelansichten 93 Artikel-Downloads Metriken
1493 ANSICHTEN 93 HERUNTERLADEN 2 Crossref ZITATE Google
Scholar
ZITATE

Artikel mit ähnlichem Inhalt:

Finite Element Analysis of Stress and Strain in Splinted vs. Non-splinted Implant Supported Mandibular Overdentures Journal of Long-Term Effects of Medical Implants, Vol.28, 2018, issue 4
Paniz Fasih
Evaluation of Stress and Strain Magnitude in Tapered and Cylindrical Implant Surrounding Bone, Used for Restoration of Mandibular Teeth: Finite Element Analysis in All-on-Four Concept Journal of Long-Term Effects of Medical Implants, Vol.31, 2021, issue 4
Mehdi Ekhlasmand Kermani, Farhad Hajizadeh, Amir Mahmoudi Motlagh, Mohammad Sadafsaz
Comparison of the Effects of Different Implant Apical Designs on the Magnitude and Distribution of Stress and Strain in Bone: A Finite Element Analysis Study Journal of Long-Term Effects of Medical Implants, Vol.24, 2014, issue 2-3
Hasti Torabi, Mohammad Reza Movahhedy, Ardeshir Lafzi, Saeed Raoofi, Mahdi Kadkhodazadeh, Maryam Khademi, Reza Amid
RETRACTED
Finite-Element Analysis of Stress and Strain in Mandibular Overdentures Supported by Splinted versus Nonsplinted Implants
Journal of Long-Term Effects of Medical Implants, Vol.28, 2018, issue 4
Amir Mahmoudi Motlagh, Majid Sedaghat Monfared, Ardeshir Lafzi, Mohammad Dehghan, Paniz Fasih, Forouzan Faghani
Effect of Implant Design on Stress Distribution: A Finite Element Study Journal of Long-Term Effects of Medical Implants, Vol.32, 2022, issue 4
Mohammadreza Talebi Ardakani, Behzad Houshmand, Iman Zoljanahi Oskoui, Anahita Moscowchi

Zukünftige Artikel

Transfer of guided emergence profile developed using customized provisional implant restoration and cementation using an abutment replica technique- A case report. Vinay Sivaswamy, Balaji Anand S D, Jacob Mathew Philip, C J Venkatakrishnan, Narasimman M, Helen Mary Abraham, Pavithra B Marijuana use is Associated with Increased Rates of Hip Dislocation and Lower Insurance Reimbursement among THA Recipients Andrew Recker, Chukwuweike Gwam, Erin Kelly, Scott Douglas, Johannes Plate Surface modification of Mg alloy AZ91D using nanopowder mixed EDM for Biodegradable Implant Alok Kumar, Abhishek Singh The history of orthopedics in the long term: from general medicine to medical specialties Pavlos Altsitzioglou, Ioannis Zafeiris, Fotini Soucacos, Pantelis Limneos, Eleftheria Souliotis, Evanthia Mitsiokapa, Theodosis Saranteas, Andreas Mavrogenis COMPARISON OF ROUGHNESS, WETTABILITY AND SCANNING ELECTRON MICROSCOPY FEATURES BETWEEN SANDBLASTED ACID-ETCHED AND OXIDIZED TITANIUM DENTAL IMPLANTS Kshitiz Chhabra, Arvina Rajasekar CORRELATION BETWEEN PRIMARY STABILITY OF DENTAL IMPLANTS AND BONE DENSITY: A RETROSPECTIVE ANALYSIS Padmaja S, Arvina Rajasekar C-Reactive Protein and Erythrocyte Sedimentation Rates after total and unicompartmental knee arthroplasty - less implant equals quicker normalization Kasım Kılıçarslan, Ömer Faruk Naldöven, Enejd Veizi, Şahan Güven, Şahin Çepni, Ahmet Fırat COMPLICATIONS FOLLOWING SHORT FEMORAL NAIL FIXATION FOR INTETROCHANTERIC HIP FRACTURES. A RETROSPECTIVE STUDY Anastasia Vasilopoulou, Antonios Trichonas, Konstantinos Palaiologos, Emmanouel Antonogiannakis, Christos Nikakis, Stavros Angelis, Spyridon Maris, Christos Grekas , Efthimios Karadimas, Alexandros Apostolopoulos FEA: Connector Designs and Pontic Stress Distribution of Fixed Partial Denture Implant-Supported Metal Framework Saja Muhsin, Enas Mohammed, Khalid Bander Design and Frequency Stability Analysis of an Adaptive Neuro-Fuzzy Inference System based Artificial Pacemaker Controller in MATLAB Nader Jafarnia Dabanloo, Asghar Dabiriaghdam, Fereidoon Nooshirvan Rahatabad, Keivan Maghooli Complications following intraosseous injections of calcium phosphate bone cement in Subchondroplasty. Christos Zachariadis, Helen Leligou, Stauros Kourkoulis, Evaggelos Magnisalis, Panayiotis Papagelopoulos, Olga Savvidou The Impact of Patient Characteristics on Outcomes of Surgically Managed Vertebral Osteomyelitis in the United States: Insights from a National Database Study Jennifer Kurowicki, Stuart Changoor, Daniel Coban, Neil Patel, Kumar Sinha, Ki Hwang, Arash Emami ASSOCIATION OF AGE, GENDER AND SITE OF IMPLANT PLACEMENT - AN INSTITUTION BASED RETROSPECTIVE STUDY Harini Sri, Nabeel Ahmed, Keerthi Shasanka
Digitales Portal Digitale Bibliothek eBooks Zeitschriften Referenzen und Berichte Forschungssammlungen Preise und Aborichtlinien Begell House Kontakt Language English 中文 Русский Português German French Spain