Very Large Diameter Polymer Acetabular Liners Show Promising Wear Simulator Results

James W. Pritchett
Hansjörg Wyss Hip and Pelvis Center, Swedish Medical Center, Seattle, WA; Orthopaedics International, 901 Boren Avenue #711, Seattle, WA 98104

ABSTRACT

Thinner and larger acetabular components are more stable and are useful for resurfacing. This study performed wear simulator tests of the largest polyurethane (PUR) and highly cross-linked polyethylene (XLPE) liners available. The results were compared with conventional polyethylene (ConvPE) to determine wear, survivorship, wear debris particles generated, and oxidative degradation.
Two-piece acetabular components with a 4-mm poly liner, a 2-mm metal backing, and 1 mm of porous coating with an inner diameter of 51 mm were tested. Cobalt chromium (CoCr) and titanium nitride (TiN)-coated titanium- resurfacing femoral components were counter-face surfaces. Acetabular components were tested at 45° and 65° inclinations, with both 300 μm and an increase of 1 mm in radial clearance.
After 30 million cycles, PUR had 56% (p < 0.001) less wear and cross-linked had 93% (p < 0.001) less wear than ConvPE. TiN-coated femoral implants had 23% less wear than CoCr when used with cross-linked polyethylene. Wear increased by 16% at 65° and by 19% when radial clearance increased with cross-linked polyethylene. Polymethylmethacrylate debris ultimately failed ConvPE, but not the cross-linked polyethylene or PUR. XLPE acetabular liners are compatible with >20 years of expected use in active patients.


Articles with similar content:

Options for Acetabular Fixation Surfaces
Journal of Long-Term Effects of Medical Implants, Vol.17, 2007, issue 3
Trevor G. Murray, Wael K. Barsoum, Hussein Darwiche, Alison K. Klika
Mechanical Stability of Polyethylene Liners Cemented Into Acetabular Shells
Critical Reviews™ in Biomedical Engineering, Vol.28, 2000, issue 1&2
Kevin D. Weaver, Christopher V. Bensen, H. Del Schutte, Jr.
Implant Debris Particle Size Affects Serum Protein Adsorption Which May Contribute to Particle Size-Based Bioreactivity Differences
Journal of Long-Term Effects of Medical Implants, Vol.24, 2014, issue 1
Anand Reddy, Marco S. Caicedo, Joshua J. Jacobs, Nadim James Hallab, Lauryn Samelko
Clinical, Surface Damage and Oxidative Performance of Poly II Tibial Inserts After Long-Term Implantation
Journal of Long-Term Effects of Medical Implants, Vol.18, 2008, issue 2
Matthew Austin, Clare M. Rimnac, Gregg R. Klein, Matthew Kraay, Peter Sharkey, Francisco Medel, Harlan Levine, Steven M. Kurtz
PYROELECTRICITY: ITS PHENOMENOLOGY AND ITS APPLICATION TO TEMPERATURE MEASUREMENT
International Heat Transfer Conference 4, Vol.35, 1970, issue
Sidney B. Lang